Customization: | Available |
---|---|
Type: | Motion Simulator |
Output Signal Type: | Digital Output |
Still deciding? Get samples of $ !
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Indicator name | Indicator parameters |
Maximum load weight | 50kg |
Table size | Φ580mm |
Mesa material | Stainless steel |
Countertop flatness | 0.02mm |
Countertop end runout | 0.03mm |
Mesa leakage | 0.2mT |
Perpendicularity of the two axes | ±5" |
The degree of intersection of two axes | Φ0.25mm ball inside |
User conductive slip ring | 60~120 rings (rated current not less than 3A) |
Communication interface | RS232/RS422 |
Rate resolution | 0.001°/s |
Angular rate accuracy and rate smoothness |
ω<1°/s, 3×10^-3 (1°average) 1°/s≤ω<10°/s, 3×10^-4 (10°average) ω≥10°/s, 3×10^-5 (360°average) |
Corner range | 360°continuous rotation |
Inside frame | |
Shafting rotation accuracy | ±2" |
Angular position measurement resolution | 0.36" |
Angular position positioning accuracy | ±5" |
Angular positioning repeatability | ±3" |
Speed range | ±0.001°/s~±500°/s |
Maximum angular acceleration | 300°/s² (rated load) |
Outer frame | |
Shaft system rotation accuracy | ±5" |
Angular position measurement resolution | 0.36" |
Angular position positioning accuracy | ±5" |
Angular positioning repeatability | ±3" |
Speed range | ±0.001°/s ~ ±200°/s |
Maximum angular acceleration | 100°/s² (rated load) |
Product name | High and low temperature test chamber |
Nominal volume (optional) | 125L~275L |
Test chamber size (WXDXH) (optional) | 500×500×~650×650×650mm |
Test chamber outer dimension (WXDXH) (optional) | Approx. : 740*95*950 mm~880×1100×1100mm |
Temperature range | -60°ºC~+90 °ºC |
Temperature control range | -60~+90 ºC |
Temperature fluctuation (optional) | <±0.5ºC~±1ºC |
Temperature uniformity | <1.0ºC |
Temperature deviation | <±2.0ºC |
Cooling rate | >2°C/min |
Heating rate | >3°C/min |
Total installed power | Approx. 13KW |
Weight of the whole machine (box + unit) | Approx. 210kg to 260kg (based on the weight after final confirmation of size) |
Cooling method | Air cooling |
Refrigeration method | Mechanical cooling provided by cascade system |
Refrigerant | Environmentally friendly and highly efficient refrigerants (R404a and R23) |
Liner material | SUS304 |
Refrigeration compressor | Two advanced compressors |
Heating system | Nichrome electric heater with PID adjustment |
Regulation mode | PID Balance |
Temperature and Humidity Control | BTHC |
Control system | Programmable controller + touch screen |
Remote control | RS485 interface, can communicate with the computer, remote monitoring; And with USB interface |
Test Port | One O25mm test hole |
Environmental requirements | Ambient temperature: 0ºC~+30ºC, relative humidity: ≤85%RH, no strong vibration, electromagnetic radiation shielding, no dust and corrosive substances around |
Power supply system | 380±10% VAC, 50Hz+N+G, ground resistance: <4Ω |
Safety Measures | Independent ultra-high and ultra-low temperature alarms, A ircraft overheating alarm, refrigeration compressor overpressure/overload alarm; Power supply missing phase, phase sequence alarm; Leakage, short circuit protection. |